Paper: | BIO-P1.8 |
Session: | Biomedical Signal Processing I |
Time: | Tuesday, May 16, 10:30 - 12:30 |
Presentation: |
Poster
|
Topic: |
Bio Imaging and Signal Processing: Biomedical signal processing |
Title: |
FREQUENCY AND TIME-FREQUENCY BASED INDICES FOR SUCTION DETECTION IN ROTARY BLOOD PUMPS |
Authors: |
Antonio Ferreira, Marwan A. Simaan, J. Robert Boston, University of Pittsburgh, United States; James F. Antaki, Carnegie Mellon University, United States |
Abstract: |
A new suction detection system for rotary blood pumps used in Left Ventricular Assist Devices is presented. The system is based on frequency indices combined with a time-frequency extraction feature algorithm. The frequency based indices can detect the changes in the harmonic and subharmonic energy content of the pump flow signal, when a suction event is occurring. The time-frequency extraction feature algorithm can track variations in the standard deviation of the instantaneous frequency of that signal. These two pieces of information are then combined in a weighted decision system to generate a suction alarm. The proposed system has been tested in simulations, in a mock-loop system, and in-vivo tests and produced very satisfactory results. |